Newsletter #11

Temperature Measurements

With the release of the TC1000, KinExA users are now able to run experiments at temperatures ranging from 4°C to 40°C. While useful for characterizing biological interactions, there are some things to consider when changing temperature.

Equilibrium & Kd shifts

Samples need to equilibrate at the temperature at which they are being evaluated. This means that samples prepared on the bench need to reach the temperature inside the TC1000 and then reach binding equilibrium at that temperature. Small samples reach temperature more quickly than large samples. Make sure to give samples enough time to reach temperature in addition to equilibration time when calculating incubation time needed.

Kd values are temperature dependent. We have routinely seen Kd’s get weaker with higher temperatures but that is not always the case. If you have a system that you measured at room temperature, do not expect the same results at other temperatures. If the Kd becomes tighter, constant binding partner (CBP) and titrant concentrations may need to shift down to accurately define the Kd. Conversely, if the Kd becomes weaker, you may need to use higher concentrations. Note that CBP concentrations need to stay in linear range and if working with weaker Kd’s, make sure you are still in the KinExA Mode. Refer to TN224 Linear Range and TN221 KinExA Mode for more information.

Condensation

Switching from 4°C to warmer temperatures can cause condensation in the instrument. Normally, if the TC1000 remains closed until the new temperature is reached, condensation will not be an issue. Note: the buffer in the reservoir will take much longer to reach the new temperature. To ensure condensation does not affect the instrument, and that buffer is at the right temperature, change the temperature at the end of the day. The instrument and buffer will be ready the next morning.

Evaporation

Small Volume Tubes (P/N: 111642) are often used for small sample volumes and to reduce dead volume needed. According to TN206 Minimum Sample Volume, as little as 10 μL is sufficient for dead volume for an experiment duration of 5 hours at room temperature. That will not be the case at 37°C; evaporation happens more quickly so additional volume should be added. While we recommend a minimum of 50 μL dead volume, less can be evaluated and potentially used.

Samples should be covered during equilibration. This is true at any temperature. Foil Tape (P/N: JVCCAF20) or scotch tape are good options, but do not use Parafilm. When the experiment starts, the sample tip will pierce the tape to access the sample. As the experiment progresses, there is a hole that allows sample to evaporate. If that is a concern due to small samples, consider preparing the titration series and then splitting them in half to aliquot into additional tubes. Cover all samples and adjust the sample set to sample each tube once.

Temperature Tolerant Parts

The XP and XCalibur Pumps used on 3X00 and 4000 KinExA models come standard with plastic XP or XCalibur Valves. In order to avoid the temperature effects of plastic pieces, Ceramic XP Valves (P/N: 310136) or Ceramic XCalibur Valves (P/N: 310131) will need to be used. Syringe barrels that contain a plastic piston should also be replaced with Temperature Tolerant Syringe Barrels (P/N: 310510). It is important that these Valves are in good working condition to prevent leaks that may lead to bubble and sample draw issues.

Note: These parts are included with the purchase of a TC1000.

TC1000 with KinExA 4000

Tube Rotator

Our tube rotator was designed to optimize KinExA experiment preparation. It employs gentle and effective mixing with an adjustable angle that optimizes mixing for any sample. The compact design is small enough to fit on a crowded bench, in an incubator, or under the hood. Purchase includes the base unit, power cord, one tube roller mixing rotor, and one end-over-end mixing rotor. Rotors can hold up to ten 50 mL tubes, ten 5 or 15 mL tubes, and up to 30 micro volume tubes.


Tips & Tricks

Aspiration Pump Syringe Purge

Sapidyne’s latest software release, KinExA Pro version 4.5, includes a new feature called the Aspiration Pump Syringe Purge. As the name suggests, this feature allows you to “purge” your instrument’s syringe barrel of those little bubbles that sometimes cling to the top of the syringe. Keeping air out of the system helps in getting clean and accurate data.

To purge the syringe open the Instrument tab, click on the “Syringe Purge” button, then click “Start”. If you are using a 3200 model, the software will ask you to loosen the Luer cap on the degasser module until the vacuum pump engages, then retighten the cap (see Figure 1). The KinExA 4000 can engage the vacuum pump through software alone, but the 3200 needs you to release some pressure to engage the vacuum pump. Afterwards, both models will automatically perform the syringe purge and inform you when it is complete (about 2 minutes). This feature only exists on KinExA Pro software version 4.5 and newer.