If contamination is suspected, use this procedure to remove residual material from the Instrument Lines. This procedure is also recommended once a month to keep the Instrument clean.
Before cleaning the Instrument, thePressure Transducerneeds to be bypassed to avoid corrosion. See How to Guide 238 – Bypassing the Pressure Transducer (HG238) for instructions.
Note:The 4000 and other Instruments with a stainless steelPressure Transducerdo not require thePressure Transducerto be bypassed.
Note:For 3X00 the various cleaning solutions should be placed on theBuffer Line,Inject Line,Sample Lines, and in theParticle Reservoir.
Whenever a new solution is introduced, aBuffer Changeprocedure should be performed to put an air bubble between the solutions. Introducing a bubble to theBuffer Linereduces mixing of the new solution with the old.
Click on theBuffer Changeicon and follow the directions provided.
Fill/EmptytheInjection Syringe[2] times in order to prime theSyringewith each new solution.
Extreme Clean Procedure
Run [10]Rinseswith a bleach solution*. WipeSample Lineswith the bleach solution.
Be sure to replace or sterilize all containers touching instrument fluids, including theBuffer Reservoir,Bead Vial, etc.
* It is important to keep the bleach solution below 0.5% NaOCl (sodium hypochlorite). If using household bleach, (~5% NaOCl) a 10 fold dilution is adequate. If using concentrated sodium hypochlorite solution (~10-15% NaOCl) a 20-30 fold dilution should be used.