The following cleaning procedure is recommended monthly or as needed to control contamination. Be sure to clean the Buffer Bottle, Bead Bottles, and the Instrument to avoid re-contamination.
KinExA 4000
Whenever a new solution is introduced, aBuffer Changeprocedure should be performed to put an air bubble between the solutions. Introducing a bubble to the Buffer Line reduces mixing of the new solution with the old. Newer versions of the KinExA Pro Software have an automated tool that guides the user through the procedure.
Click on theBuffer Changeicon and follow the directions provided.
Bleach Solution*:Open theFast Rinsescreen and run [10] cycles.
Wash Station
First remove the plumbing connections then remove the Wash Station. Some Wash Stations on the KinExA 4000’s have a screw at the bottom of the Wash Station which must be removed using a T10 driver before the Wash Station can slide out. Other Wash Stations are removed simply by turning it counter-clockwise (left). Be careful when removing the plumbing from the Wash Station as some liquid will run out.
KinExA® Cleaning Solution:OpenFile > New from Template > Cleaning > Standard Wash Template. Add45 mLof cleaning solution to Standards Rack: Tube 1. Open theTiming Setupand adjust the soak time by changing the time for the buffer step following Standards Tube 1 to 1200 seconds. Change the number of cycles to [10] and pressStart.
Note:A full 250 mL bottle of cleaning solution is needed.
Repeat theBuffer Changeprocedure to switch to a buffer of choice.
Buffer of choice:Open up theFast Rinsescreen and run [10] cycles.
* It is important to keep the bleach solution below 0.5% NaOCl (sodium hypochlorite). If using household bleach, (5% NaOCl) a 10 fold dilution is adequate. If using concentrated sodium hypochlorite solution (10-15% NaOCl) a 20 to 30 fold dilution should be used.