How to Guide 211
Injection Syringe Barrel Wash
How to Guide 211 • HG211R1
Figure 1. Kinetics Inject experiment demonstrating syringe barrel contamination. The steady increase in the baseline is due to residual label in the injection syringe.
Figure 1. Kinetics Inject experiment demonstrating syringe barrel contamination. The steady increase in the baseline is due to residual label in the injection syringe.
Procedure for washing the injection syringe barrel
- Open the rinse screen.
- Fill, then empty the injection syringe with KinExA Cleaning Solution twice by selecting the fill/empty inject button. Empty the liquid to waste between rinses.
- Run two rinses with the cleaning solution, selecting only the Backflush, Inject, and Buffer.
- Fill, then empty the injection syringe with the buffer of choice twice by selecting the fill/empty inject button. Empty the liquid to waste between rinses.
- Run two rinses with the buffer of choice, selecting only the Backflush, Inject, and Buffer.
- To ensure that the injection syringe barrel is clean, run three rinses with the buffer of choice, selecting all lines. Traces should look similar to the ones seen in Figure 3B.